Reference values you can expect with the DTX-EFM2 | Fluke Networks

Base de conhecimento

KB Home   >  Teste de fibra   >  Adaptadores da fibra DTX-EFM2

Reference values you can expect with the DTX-EFM2

Reference values give a good indication that the sources are operating as expected. They DO NOT give any assurance that the test reference cords are good. If you are reading this article because you are getting negative loss readings, you should review this article.
 
As a guide, when setting a reference, you should expect better than:

  • DTX-EFM2 62.5/125 µm: -18.5 dBm to -20.00 dBm
  • DTX-EFM2 50/125 µm: -22.0 dBm -24.50 dBm 
     
    where -23.5 dBm is better than -24.50 dBm.

Author: Adrian Young

Creation Date: 2014-03-26

Last Modified: 2014-03-26

 
 
 
 
Powered By OneLink