Error message - NO FACTORY CAL DATA FOUND

The last OMNIScanner left the factory back in July 2005. We regret to say service and support ended July 2009. It's replaced by the Versiv familly of testers.
 
The calibration EPROM on the measurement board and the OMNIScanner database calibration time stamp record contain different time stamps, so the calibration is invalid.
  
            
 
This error screen may also appear when an OMNI-LIA101 adapter with an EEProm programming error is attached to the Omni Remote. In such a case, its the OMNI-LIA101 adapter, not the OMNIScanner and Remote that caused the error screen to display. Replace the OMNI-LIA101 adapter with a good adapter.
 